Skip to Content
Reference: BV20.V01

Additional Bonnette Galilee

Public price (Tax incl.) 1,508.44 €

General conditions
Delivery: 2-3 working days

Galilean system - Principle:

  • The Galilean monocular is an afocal magnifier, meaning the image is formed at infinity.
  • The point of this system is simply to magnify for distance vision.
  • For near vision, simply add a cap lens to the front.
  • Note: its power determines the working distance, which will be at its focal length.

Monocular:

  • Body made of anodised aluminium, very stable.
  • Very light.
  • Achromatic telescope.
  • Wide field of view. 24° for the 1.8x
  • Anti-reflective coated.

Telescopes focused at infinity ∞

1.8 x

2.1 x

2.5 x

Ø Objective20mm20mm20mm
Field24°20°16°
Weight10 gr12 gr16 gr
  • You can:
    • Buy a trial case with 3 monoculars that fit the optician's standard trial frame
    • Rent a trial case for a one-off sale.

Our add-on lens holder lets the optician find the wearer's optimal magnification using the trial box.

  • Universal: you only need one telescope with a trial ring.
  • No more need for huge stocks.
  • Placed on the trial frame, it lets you add the patient's prescription to assess the benefit of correcting.
  • The trial is done in real conditions.
    • With a reading chart or a newspaper.
    • On a computer screen.
    • For intermediate vision, a TV screen.
  • To find the add-on lens power, divide 100 by the eye/object distance (in centimetres).
  • Choose a large "panto"-style frame.
    • The optical centre must be the near vision centre.
    • The optical centre-to-frame distance must be at least 19mm.
    • Mark the optical centre (near vision) and check it.
  • For a monocular fitting, order:
  1. The chosen 1.8x, 2.1x or 2.5x telescope (Code BV20.Gxx ?).
  2. The add-on lens determined with the trial box. Code (BV20.Axx ?)
  • Option: pre-sized ophthalmic lens, to be edged to 21mm.
  • The plain mounting ring (Code BV20.B24) or prescription mounting ring (Code BV20.B20), depending on the option above
  • A mounting disc to be edged. (Code BV20.D30)
  • A large case. (Code BV20.E01)
  • For a binocular fitting, order:
    1. 2 chosen 1.8x, 2.1x or 2.5x telescopes (Code BV20.Gxx ?).
    2. 2 add-on lenses determined with the trial box. Code (BV20.Axx ?)
    • Option: 2 pre-sized ophthalmic lenses, to be edged to 21mm.
  • 2 plain mounting rings (Code BV20.B24) or prescription mounting rings (Code BV20.B20), depending on the option above
  • 2 mounting discs to be edged. (Code BV20.Dxx ?).
  • A large case. (Code BV20.E01)
  • A mounting kit. (Code BV20.D34)

    • Or simply mark the near-vision optical centre on the glasses and LESA FRANCE takes care of everything..............Code BV20.MONT
      • Simple, isn't it?........

    How to proceed

    Removable cap lens conversion table

     

    Telescope magnification

     

    1.8 x

    2.1 x

    2.5 x

    0.5 D2 metresWatching TV2.002.402.80
    1 D1 metreWatching TV2.252.603.10
    1.5 D66 centimetresLooking at a screen2.502.903.40
    2 D50 centimetresLooking at a screen2.703.103.75
    3 D33 centimetresLooking at a screen, reading3.103.604.30
    4 D25 centimetresLooking at a screen, reading3.604.205.00
    5 D25 centimetresLooking at a screen, reading4.004.705.60
    6 D17 centimetresLooking at a screen, reading4.505.206.20
    8 D12 centimetresLooking at a screen, reading5.406.307.50
    10 D10 centimetresLooking at a screen, reading6.307.308.70
    12 D8 centimetresLooking at a screen, reading7.208.4010.00
    14 D7 centimetresLooking at a screen, reading8.109.4011.20
    16 D6 centimetresLooking at a screen, reading9.0010.5012.50

    How to choose the telescope magnification?


    • Look at the table above.
    1. Suppose the patient has useful acuity with a Galilée 1.8x and a 12D cap lens: the working distance is 8cm.
    2. With a 2.1x telescope, the working distance goes from 8cm to 10cm for the same magnification (±≈).
    3. With a 2.5x telescope, the working distance goes from 8cm to 12cm for the same magnification (±≈).
  • You gain distance on one side but lose field on the other; testing is needed to find the best solution.


  • The Galilean telescope.

     
     
    Telescopic systems are multi-lens systems that allow both distance and near vision.
     
     

    How does it work?

    Its basic principle is that of binoculars.

     

    Binoculars are a Keplerian system.

     
    • It consists of a convex objective and a concave eyepiece..
    • When the 2 focal points coincide, the system is said to be afocal.
    • The magnification of such a system is the ratio of the objective's focal length to the eyepiece's.
    • So with an objective of 80mm and an eyepiece of 40mm focal length, the magnification is 80/40=2.
    • The total length of the system is then 120mm, and the image is also inverted.
    • Rather hard to mount on spectacles.

    The Galilean telescope is the most convincing alternative.

    • It consists of a convex objective and a concave eyepiece.
    • When the 2 focal points coincide, the system is said to be afocal.
    • the 2 focal points coincide; since the eyepiece is negative, both lenses are in front of the focal point.
    • The total length of the system is then 40mm and the image is upright, so no erecting prisms are needed.
    • The drawback is that the eyepiece focal length is very long for a negative lens, which limits magnification to 3x at most to keep a good image.
    • That said, the system is compact and light, so it can be mounted on a spectacle frame.

    Vision up close.

    • Adding a lens in front of a telescope makes the system's focal length that of the add-on lens.
    • The magnification is the telescope power multiplied by that of the lens.
    • For binocular vision, angled mounts are used.
    • The limit for a binocular fitting is 5D.
    • To correct ametropia, a prescription lens is placed in front of the eyepiece.
    • Our company can carry out the complete fitting.

    In summary.

    • The system is light and can be mounted on a frame.
    • A Galilean telescope without a lens gives vision at infinity with corrected ametropia
    • Near vision at the focal length of the add-on lens.
    • Without corrected ametropia the image is still fine; only the focal distance varies up or down.
    • High astigmatism must be corrected.
    • To correct ametropia, a prescription lens is placed in front of the eyepiece.
    • By design, our telescopes have a pinhole effect.

    The big advantage.

    • Allows hands-free work.
    • Allows watching TV with additions below 1.
    • Lets you read a whole screen just by moving your head.

    ▲▲▲▲▲Top of page▲▲▲▲

    Characteristics:

    Reference BV20.V01
    Public price (Tax incl.) 1,508.44 €

    Content:

    Qty
    [BV20.G18] Monocular of Galilean (1.8 x) 1
    [BV20.G21] Monocular of Galilean (2.1 x) 1
    [BV20.G25] Monocular of Galilean (2.5 x) 1
    [BV20.B24] Tightening ring (Without correction) 3
    [BV20.D36] Mounting disc for test glasses 3
    [BV20.U] Additional Bonnette Galilee 1